TezSafety Series Class2 A2 Microbiological Safety Cabinets
Class 2 Microbiological Safety Cabinets (MSC) are devices that provide high security, protecting employees, the environment and their products. The design of these cabinets allows for the safe intake of biological agents through the correct direction and filtration of air. Airflow and Filtration: Class 2 Biological Safety Cabinet, the flexibility of operation begins with air being drawn in and this air is stored in the HEPA filters at the bottom of the cabinet and then thrown out. This arrangement directs the operator, product and environment through the air in the form of particles. The ventilation system ensures that 70% of the air is circulated in the cabinet, while 30% is equipped as filtered exhaust air.
Tez Safety Series CLASS II A2 Cabinet TECHNICAL SPECIFICATIONS
The biologically contaminated plenum is under negative pressure or surrounded by negative pressures.
Air sources: %30 Exhaust %70 Internal Circulation
Control Electronic Control System
HEPA Filters 2 Units Production in accordance with 12469 Standards
Control panel has filter life hours, UV times hours, window opening and closing, fan controls, fan efficiency, front camera barriers condition, all alarms, UV countdown timing, device total operating time
TSK-90 External Dimensions: 1050 x850x 2200 mm Internal Dimensions: 900 x 630 x 625 mm LCD Controls
TSK-120 External Dimensions: 1340x850x 2200 mm Internal Dimensions: 1190 x 630 x 625 mm LCD Controls
TSK-150 External Dimensions: 1640 x850x 2200 mm Internal Dimensions: 1490 x 630 x 625 mm LCD Controller
Tezsafety TSK-120 MGK Cabin Technical Specifications
Internal Dimensions (WxLxH) mm 1190x630x625
External Body Dimensions (WxLxH) mm 1340x850x1450
Carrier Stand Dimensions (WxLxH) mm 1340x850x745
Working Floor Height from Ground 840 mm
Air Flow Rate (m/s) Inlet 0.50 ± %5 Down Flow 0.38 ± %5
HEPA Filter Type (EN 1822)
Main Filter H14 HEPA 0.3 µm %99.995
Exhaust Filter H14 HEPA 0.3 µm %99.995
Pre-Filter (EN779, Optional) EU4
Working Area Class EN ISO 14644-3 < ISO 5 US FED 209E < Class 100
Working Floor Standard 304 Stainless Steel
Optional 316L Stainless Steel
Windshield (mm) Preset/Std 200cm ± 10mm Max 530cm ± 10mm
Noise Level Normal Mode < 56 dB(A) Eco (Standby) Mode < 50 dB(A)
Working Area (Lux) 850 – 1250 Lux Frontshield (mm) < 6 mm
Energy Consumption (kW) Fans 200 W
Lighting 50 W UV Lamp 30 W Sockets
220-240 VAC / 16A Voltage / Frequency / Power / Current 230 VAC / 50 Hz / 2500 W / 16A Packaging /
Palletized Dimensions (WxLxH) mm 1645x980x1840
Weight (kg), Body 300 kg, excluding carrier stand, kilograms.